The Group3 High Field Hall Effect Probe is most suitable to be use with a DTM-151 or DTM-152 Digital Teslameter and, depending on your order, is capable of measuring magnetic field up to 13 Tesla or 130,000 Gauss.
Accuracy at 25°C:
- ±0.05% of reading + 0.006% of full scale above 2.2T
Operating Range:
Up to 6T
4- Range Operation
0.3, 0.6, 1.2, 6.0 Tesla Full Scale
3, 6, 12, 60 Kilo Gauss Full Scale
Up to 8T
4- Range Operation
0.3, 0.6, 1.2, 8.0 Tesla Full Scale
3, 6, 12, 80 Kilo Gauss Full Scale
From 9 to 13T
4- Range Operation (only top range capable to read up to 13T)
1.2, 2.2, 2.2, 13.0 Tesla Full Scale
12, 22, 22, 130 Kilo Gauss Full Scale
Temperature Stability:
- Calibration: ±10ppm of reading/°C max.
- - 3ppm/°C of reading per meter of probe cable
- Zero Drift: ±(1μT + 0.0003% of full scale)/°C max.
Operating Temperature Range:
- 0 to 50°C operating to spec, -20 to +60°C max.
Dimensions:
- Probe Head Size: 14 x 5 x 2 mm Sensitive Area: 1 x 0.5 mm
- Unshielded part of cable at probe head: 5.0 ±0.2mm, 300 mm nominal length
- Shielded Cable: Ø 6.5 ±0.2 mm
NOTE:
This probe is intended for high field measurements above 2.2T up to its maximum calibrated field point. Use a standard MPT-141 probe if measuring fields below 2.2T.